VRO Encoder Readout Troubleshooting Guide
Troubleshooting

VRO Encoder Readout Troubleshooting Guide

VRO Readout Troubleshooting

First - always read the Users Manual.

SymptomPossible Cause
The carriage is moving but the Readout is not changing.Check the plugs to ensure they are connected properly.
 Check for any loose wires.
 The reading head may be too far from the encoder. Make sure they are in close proximity but not touching.
Readout is showing the wrong unit of measure.Check to make sure you are using the right encoder settings. Reset the scale. (All the encoders available from Velmex are programmed into the VRO Readout.).
The display is blank but light is flashing.VRO has been put in sleep mode. To exit sleep press "U" button, or send any character in the Serial port.
Status light flashing rapidly and display shows partial information or odd characters.Power is intermittent or was interrupted. Check power input connector and cycle power.
Status light pulsating and display shows ">5 INPUT VOLTS !"Power in is greater than 5.4 volts, disconnect power adapter and check the voltage.
 Power in is less than 4.6 volts, check power adapter voltage, and encoder load.
Status light pulsating and display shows "Enc Input Fail !"Poor encoder connection, electrical interference, or count exceeding 1.6 MHz. command.
It will not let me make setting changes. Displays "Not Settable".VRO needs to be cleared (set to defaults) and the process restarted.

 

×
×
×
×
×